Abstract

This work describes a new method for the correction of signal drift of a MS-based sensor by standard gas addition (SGA). It consists of introducing a gaseous He-Xe mixture continuously and independently of the carrier gas into the mass spectrometer source. To mimic the disturbances generated by periodical tuning of the mass detector, three series of adjustments were made to the main acquisition parameters. The influence of the SGA signal correction on the discriminating power of the data was evaluated from the analysis of three batches of cheeses by dynamic headspace coupled to mass spectrometry. SGA afforded a good correction of the main types of drift classically observed in mass spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.