Abstract

Accuracy is paramount in evaluating sperm concentration and can be a challenge to those with minimal laboratory experience. Purpose of the study was to determine the effect of operator experience on sperm concentrations using 4 methods: Makler® counting chamber, equine densimeter, iSperm, and NucleoCounter®. There was no difference (p = 0.64) between experienced and novice processors for Makler® counting chamber; was difference (p = 0.005) for equine densimeter; was no difference (p = 0.35) for iSperm; and was tending toward difference (p = 0.068) for NucleoCounter®. Correlation between bias and magnitude for Makler® counting chamber was –0.74 (p = 0.003); for equine densimeter was –0.44 (p = 0.11); for iSperm was 0.06 (p = 0.83), and NucleoCounter® was –0.52 (p = 0.06). Makler® counting chamber produced a mean sperm concentration similar to an experienced processor but had significant variation within novice processors. Equine densimeter significantly overestimated sperm concentrations with novice processors but had the least variation. The iSperm performed poorly for both experienced and novices and produced significantly different concentrations than other methods. Thus, iSperm cannot be recommended to accurately measure sperm concentration. Lastly, NucleoCounter had no difference in mean sperm concentrations or variation and was the best system for a novice processor to gain accurate and repeatable sperm concentration measurements.