Abstract

We compared the accuracy of three common methods of total protein normalization. The Stain-Free method was accurate across different types/brands of western blotting membrane and for various protein loads, unlike Ponceau S and Amido Black. Normalizing to the housekeeping proteins Actin and β-Tubulin could match the accuracy of the Stain-Free method. However, compared to Actin or β-Tubulin, normalizing to the Stain-Free signal reduced variability that led to enhanced reproducibility and a reduction in the number of samples needed to obtain statistically significant results by >50%. Stain-Free normalization can enhance the reproducibility and hence the confidence in Western Blot data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.