Abstract

Introduction. Transcranial magnetic stimulation or repetitive transcranial magnetic stimulation (TMS/rTMS) are currently used in research and treatments of diseases of the central nervous system, such as recurring depression. Strong electric pulses are used to produce strong pulsed magnetic fields that are directed to the patient’s cerebral cortex where the fields induce electric pulses. The pulses may be causing unnecessary exposure of the staff. Method. The MagVenture TMS/rTMS system was investigated, without patient presence, through measurements of magnetic field pulses at varying distances from the emitting coil and different power settings (94–127 A/s). Results. Fourteen measurements were done which displayed exposures exceeding the given guidelines up until a distance of 40 cm from the transmitting coil. Discussion. The study shows that the exposure of staff in this type of treatment may exceed the given guidelines for occupational exposure, thus confirming previous findings. This necessitates good routines in information and treatment procedures to avoid this exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.