Abstract

The classical CPMG sequence and its extension as an imaging sequence, fast spin echo (FSE, based on RARE), suffer from signal magnitude variations in the early echoes when the refocusing pulses are not set exactly to 180 degrees. It has been suggested that by varying the value of the nutation angle of each refocusing pulse the signal magnitude could be made constant. This article describes an algorithm permitting the generation of sequences of nutation angles yielding series of echoes with constant signal magnitudes. This result is then used to design selective pulses for the FSE imaging technique.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.