Abstract
To help preserve in-theater strength within deployed military units, commercially available, rapid, user-friendly ABO-Rh blood typing kits were evaluated to determine their stability in storage conditions commonly encountered by the warfighter. Methods for environmental exposure testing were based on MIL-STD-810F. When Eldon Home Kits 2511 were exposed to various temperature/relative humidity conditions, the results were comparable to those obtained with the control group and those obtained with industry-standard methods. For the ABO-Rh Combination Blood Typing Experiment Kits, 2 of the exposure treatments rendered them unusable. In addition, a third set of exposure treatments adversely affected the kits, resulting in approximately 30% blood type misclassifications. Collectively, this evaluation of commercial blood typing kits revealed that diagnostic performance can vary between products, lots, and environmental storage conditions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.