Abstract

VACUTAINER PPT plasma preparation tubes were evaluated to determine the effects of various handling and shipping conditions on plasma human immunodeficiency virus (HIV) load determinations. Plasmas obtained from PPT tubes stored and shipped under nine different conditions were compared to conventional EDTA tube plasmas stored at -70 degrees C within 2 h after phlebotomy. Compared to viral loads in frozen EDTA plasma, those in PPT tube plasma that was frozen immediately and either separated or shipped in situ were not significantly different. Viral loads in PPT tube plasma after storage for 6 h at either room temperature or 4 degrees C, followed by shipment at ambient temperature or on wet or dry ice, were not significantly different from baseline viral loads in EDTA or PPT plasma. The results of this study indicate that the HIV load in PPT tube plasma is equivalent to that in standard EDTA plasma. Plasma viral load is not affected by storage or shipment temperature when plasma is collected in PPT tubes. Furthermore, plasmas can be shipped in spun PPT tubes, and the tubes provide a safer and more convenient method for sample collection and transport than regular EDTA tubes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.