Abstract

ObjectivesOur study aimed to evaluate the stability of human immunodeficiency virus 1 (HIV-1) RNA on cobas plasma separation card (PSC) specimens for viral load (VL) testing after being exposed to varied temperatures and storage times.MethodsFor this purpose, venous PSC specimens were collected and stored at 25ºC to 42ºC for a period of up to 28 days. Plasma VL at baseline was used as reference, against which PSC VL was compared at different time points.ResultsFrom the 30 patients included in the study, 600 PSC and 30 fresh plasma specimens were obtained. Plasma VL at baseline was fewer than 1,000 copies/mL in 16 patients, and 99.4% of PSCs from these patients yielded nonquantifiable VL at all temperature ranges and time points. During the study period, minor variation of VL was observed in PSCs obtained from 13 patients with plasma VL fewer than 1,000 copies/mL at baseline. For the patient with plasma VL at 1,000 copies/mL, the PSC VL varied from undetectable to 1,670 copies/mL.ConclusionsOur results show minor variation of VL in PSC specimens in the study conditions. HIV RNA is stable in PSCs exposed to high temperatures for up to 28 days.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.