Abstract

Fixation of vertical femoral neck osteotomies in 50 cadavers was performed with either von Bahr screws or a sliding hip compression screw. One specimen from each pair of femora was used for the osteotomy, the other serving as an intact control. At 0.05 r of torsion the load-deformation test showed that three von Bahr screws provided the strongest fixation, and this was confirmed by the the ultimate torsional moment test. Regardless of positioning, even two von Bahr screws were stronger than the sliding compression screw with or without an additional lag screw. The results indicate that the best torsional stability in femoral neck fractures can be obtained with three 5.5-mm screws.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.