Abstract

In this work, large area SiC BJTs with good long-term stability in 1000 hrs DC stress tests are demonstrated. It is also illustrated how wafer scanning techniques can be used to reject BJT dies with basal plane dislocations, thereby eliminating the risk for bipolar degradation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.