Abstract
Objectives We studied the stability of 33 analytes related to clinical chemistry, bone, and vitamin metabolism, after storage in serum separator tubes (SST™). Design and methods Blood was collected from 6 subjects using SST tubes. Some serum remained in the tube in contact with the barrier gel and was stored at − 80 °C for 12 mo. Results Clinically significant changes occurred only in 1,25-dihydroxyvitamin D and retinol-binding protein. Conclusions Freezing SST tubes before sample analysis is a viable option for some analytes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.