Abstract

An analytical, accurate, precise, specific, efficient and simple Ultra-Performance Liquid Chromatography method has been developed and validated for the determination of Pazopanib in bulk and was applied on marketed Pharmaceutical Dosage form. The mobile phase used for the chromatographic runs consisted of 0.1% OPA Buffer and Acetonitrile in the ratio of 30:70% v/v. The separation was achieved on a BHEL UPLC column using isocratic mode. Pazopanib Drug peak were well separated and were detected by a PDA detector at 256 nm. The developed method was linear at the concentration range 6-14 μg/ml for Pazopanib. The method has been validated according to ICH guidelines with respect to system suitability, specificity, precision, accuracy and robustness. The LOD and LOQ for the Pazopanib were found to be 0.5853 µg/ml and 1.7738µg/ml respectively. The developed method is simple, precise, specific, accurate and rapid, making it suitable for estimation of Pazopanib in bulk and marketed pharmaceutical dosage form dosage form.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.