Abstract
In the present study, a high-performance liquid chromatographic method was validated to evaluate purity of Sertraline Hydrochloride in tablet dosage form. Sertraline Hydrochloride and two related substances, trans isomer and 2,3-dichloro isomer of Sertraline hydrochloride, were well resolved using a ZORBAX RX C 8 column (250 mm × 4.6 mm, 5) with the mobile phase consisting of 50mM KH 2 PO 4 containing 0.1% octane sulfonic acid sodium salt (pH: 2.5 with Dil. H 3 PO 4 , 0.2% Triethylamine) and methanol (43:57, v/v). Flow rate was kept at 1.0ml/min and the detection was carried out by UV at 215nm. Resolution between any pair of isomers was found to be more than 2. The Limit of detection and Limit of quantification for both trans isomer and 2,3-dichloro isomer of Sertraline Hydrochloride were 10ng/ml and 30ng/ml, respectively. The developed and optimized method was applicable for routine determination of Sertraline Hydrochloride and its trans isomer and 2,3-dichloro isomer in tablet dosage form with sufficient selectivity, precision and accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.