Abstract

Racecadotril is an anti-diarrheal drug. Waters Alliance e2695 with PDA detector (2998) Zorbax SB C18 column (150 x 4.6 mm, 3.5 μm) (Isocratic mode) was used for the quantification of Racecadotril. A mixture of water and Acetonitrile (50:50) was used as diluent and the injection volume was 20 μl. The flow rate was 1.0mL/min (Detection wave length 230 nm). Linearity was evaluated in the concentration range 1–360 μg/ml with regression equation y = 14710 x+9958.9(Correlation coefficient R2 = 0.9996). The LOD and LOQ were found to be 0.2942 μg/mL and 0.9694 μg/mL respectively. Forced degradation studies were conducted for Racecadotril capsules and method validation was performed as per ICH guidelines.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.