Abstract

Two selective stability-indicating HPLC methods are described for determination of rabeprazole sodium (RZ)–mosapride citrate (MR) and RZ–itopride hydrochloride (IO) mixtures in the presence of their ICH-stress formed degradation products. Separations were achieved on X-Bridge C18 column using two mobile phases: the first for RZ–MR mixture consisted of acetonitrile: 0.025M KH2PO4 solution: TEA (30:69:1 v/v; pH 7.0); the second for RZ–IO mixture was at ratio of 25:74:1 (v/v; pH 9.25). The detection wavelength was 283nm. The two methods were validated and validation acceptance criteria were met in all cases. Peak purity testing using contrast angle theory, relative absorbance and logA versus the wavelengths plots were presented. The % recoveries of the intact drugs were between 99.1% and 102.2% with RSD% values less than 1.6%. Application of the proposed HPLC methods indicated that the methods could be adopted to follow the stability of their formulations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.