Abstract
This study evaluated the stability and the compatibility of mixtures of morphine sulphate, bupivacaine, and clonidine hydrochloride and of hydromorphone, bupivacaine, and clonidine hydrochloride, when used in constant flow implantable pumps under simulated clinical use conditions. The pumps were filled with drug mixtures and incubated at 37 degrees C for a period of 90 days. Aliquots were sampled monthly from the reservoir and catheter outlet and the drug concentrations analysed using validated chromatography methods. Individual materials from the infusion system were immersed in the drug mixtures and stored at 37 degrees C for 60 weeks and evaluated for mechanical performance for compatibility testing. Both drug mixtures were found to be stable over 90 days in the pump at 37 degrees C. All device materials retained acceptable mechanical performance following exposure. These results demonstrate that both drug mixtures are stable when maintained at simulated body temperature in an implantable infusion system for 90 days.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.