Abstract

A technique for rapid in-plane phase-contrast imaging with high signal-to-noise ratio (SNR) is described. Velocity-encoding is achieved by oscillating the readout gradient, such that each 2DFT phase-encode is acquired three times following a single RF slice-selective excitation. Three images are reconstructed, from which both flow velocity and local resonance offset are calculated. This technique is compatible with both gradient-recalled echo (GRE) and balanced steady-state free precession (SSFP) imaging using a single steady-state. The proposed technique enables 1D velocity mapping with 40% higher temporal resolution and 80% higher SNR, compared to conventional PC-MRI using bipolar velocity-encoding gradient pulses.