Abstract

The clinical incidence of squeaking has been reported with increasing frequency, with ceramic-on-ceramic bearings seemingly most affected. This study investigated potential causes of squeaking in hard-on-hard hip bearings through 2 sets of experimental conditions. Bearing clearance appeared to affect the incidence of squeaking in metal-on-metal surface arthroplasties. The addition of third-body particles to the interface for total hip arthroplasties also affected the incidence of squeaking. In both studies, the incidence of squeaking correlated well with elevated friction. The findings of this study suggest that a likely cause of squeaking in the hip arthroplasty is adverse tribological conditions caused by suboptimal lubrication. There are numerous factors that may cause the suboptimal lubrication, and therefore, it is unlikely that an individual cause for squeaking will be identified.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.