Abstract

We describe herein a case of life-threatening hypoglycemia due to spurious elevation of glucose concentration during the administration of ascorbic acid in a type 2 diabetic patient. A 31-year-old female was admitted for proliferative diabetic retinopathy treatment and prescribed high dose ascorbic acid. During hospitalization, she suddenly lost her consciousness and her glucose concentration was 291 mg/dL, measured using self-monitoring blood glucose (SMBG) device, while venous blood glucose concentration was 12 mg/dL. After intravenous injection of 50% glucose solution, the patient became alert. We reasoned that glucose measurement by SMBG device was interfered by ascorbic acid. Physicians should be aware of this interference; high dose ascorbic acid may cause spurious elevation of glucose concentration when measuring with SMBG devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.