Abstract

Simultaneous EEG-fMRI provides an increasingly attractive research tool to investigate cognitive processes with high temporal and spatial resolution. However, artifacts in EEG data introduced by the MR scanner still remain a major obstacle. This study, employing commonly used artifact correction steps, shows that head motion, one overlooked major source of artifacts in EEG-fMRI data, can cause plausible EEG effects and EEG–BOLD correlations. Specifically, low-frequency EEG (<20Hz) is strongly correlated with in-scanner movement. Accordingly, minor head motion (<0.2mm) induces spurious effects in a twofold manner: Small differences in task-correlated motion elicit spurious low-frequency effects, and, as motion concurrently influences fMRI data, EEG–BOLD correlations closely match motion-fMRI correlations. We demonstrate these effects in a memory encoding experiment showing that obtained theta power (~3–7Hz) effects and channel-level theta–BOLD correlations reflect motion in the scanner. These findings highlight an important caveat that needs to be addressed by future EEG-fMRI studies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.