Abstract

This study was conducted to identify the device factors influencing spray pattern and particle size to gain a more complete understanding of spray plume measurements. A statistically designed experiment was used to investigate the influence of three actuator features (orifice diameter, expansion chamber depth, and orifice length) on spray pattern and particle size profiles. Custom-built actuators were manufactured and analyzed with laser light sheet illumination methods for spray patterns and laser diffraction for particle size analysis. In addition to orifice size, spray patterns were significantly influenced by the actuator orifice length and sump depth. Particle size analysis of the plumes generated from actuators used in these studies showed that all actuator features (orifice size, length, and sump depth) were significant factors influencing particle size. The performance of propellant-based metered dose inhaler aerosols seems to be significantly related to sump depth and orifice length, in addition to orifice size. Rational design of propellant-based metered dose inhalers should therefore consider these variables in addition to formulation strategies and simply modifying orifice diameter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.