Abstract

A method for precise determination of 207Pb/206Pb and 206Pb/238U ratios of zircon was developed, using an ICP-MS with 213 nm Nd-YAG laser ablation system. Compared with the conventional 266 nm laser, the 213 nm laser has advantages of high absorption of energy and effective reduction of U-Pb elemental fractionation. The maximum fractionation during the first 80 seconds is about 1.3 times the initial value. In addition, by using a linear fit of 206Pb/238U ratios acquired during a depth profile analysis, U-Pb elemental fractionation can be further corrected and the precision of 206Pb/238U ratio measurement is improved to around 3% (2σ).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.