Abstract

Changes in uric acid levels in biological fluids may be indicative of certain diseases related to kidney dysfunctions. Analytical methods recommended for the quantification of uric acid, such as high-performance liquid chromatography, are typically time-consuming and require high-cost equipment. In this study, a simple and cost-effective approach was developed for the determination of uric acid in saliva, by combining a spot test with smartphone-based digital image colorimetry. It was based on the reduction of Cu(II) to Cu(I) by uric acid, followed by complexation with 2,2′-biquinoline-4,4′-dicarboxylic acid (BCA), resulting in a violet product. Derivatization reactions and photometric measurements were both carried out in Eppendorf® tubes to facilitate the manipulation of biological samples. A smartphone camera and a free application (PhotoMetrix® 1.8) were used to capture the image and convert it to RGB values. Photometric measurements were based on reflected radiation in the G channel, corresponding to the complementary color of the product, yielding a linear response between 25 and 150 µmol L−1. The coefficient of variation (n = 10) and the limit of detection were 1.4% and 8 µmol L−1, respectively. Both reagent consumption (8.3 µg Cu(II), 3.8 mg ammonium acetate, and 156 µg BCA per determination) and waste volume (900 µL per determination) were minimized. The results agreed with those obtained by HPLC, demonstrating that the approach is a reliable alternative for large scale clinical analysis.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call