Abstract

Spot size is one of the parameters to characterize the performance of a radiographic X-ray source. It determines the degree of blurring due to magnification directly. In recent years, a variety of measurement methods have been used to diagnose X-ray spot size at Laboratory of Accelerator Physics and Application (LAPA). Computer simulations and experiments showed that using a rolled-edge to measure the spot size are more accurate, and the intensity distribution of X-ray source was obtained by a device with a square aperture. Experimental and simulation results on a flash X-ray source at our laboratory are presented and discussed in this paper. In addition, a new method for time resolved diagnostics of X-ray spot size is introduced too.