Abstract
A totally implantable subcutaneous infusion port was inserted via the left subclavian vein with peel-away sheath in a 45-year-old male patient receiving home parenteral nutrition. Sixteen months after implantation, the patient noted pain in his left clavicular region during home infusion. This was found to be due to the leakage of infusion fluid resulting from partial fracture of the catheter at the level where the catheter passed between the clavicle and first rib. Because the fracture was partial, the catheter did not embolize to the heart or large vessels. Although subcutaneous infusion ports afford patients greater freedom and the ability to lead a more active life than do external catheters, it is emphasized that there may be the risk of spontaneous catheter fracture in patients using subcutaneous infusion ports.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.