Abstract

PURPOSE: To report the clinical and histological findings of a luxated intraocular lens (IOL) in the capsular bag. METHODS: Review of a case. RESULTS: Twenty-three months after a triple procedure of vitrectomy, phacoemulsification, and IOL implantation for diabetic vitreous hemorrhage and cataract, the encapsulated IOL spontaneously luxated. Scanning electron microscopy showed sparsely distributed anterior and equatorial zonules, with only a few posterior zonules on the surface of the removed capusular bag. CONCLUSION: The absence of the anterior hyaloid membrane and posterior zonules and contraction of the lens capsule may cause dialysis of the zonules. Therefore, the anterior hyaloid membrane should be left in place in patients at low risk for the development of postoperative proliferation to maintain the long-term stability of the IOL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.