Abstract

This paper describes a complication of selective femorocerebral catheterization which was due to iatrogenic embolization of atherosclerotic material originating from the bifurcation of the common carotid artery. After analysis of the catheter maneuvers inherent to the special shape of the type of catheter used (Sidewinder, Cordis Corporation), it is concluded that this complication resulted from incorrect catheter manipulation and an incorrect choice of the dimensions of the preshaped part of the catheter in relation to the diameter of the aortic arch. Two recommendations to avoid this complication are made.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.