Abstract

Spiral imaging has a number of advantages for ultrafast data acquisition. However, implementation on high-field small-bore systems requires carefully addressing the issues of inhomogeneity-induced blurring and gradient hardware constraints. In this paper, spiral imaging on a 40-cm-bore 4.7T CSI Omega System (Bruker Instruments) is discussed. A constant-voltage gradient waveform design algorithm is developed to reduce readout times as well as minimize waveform distortions due to gradient amplifier nonlinearities. Residual errors are then measured and taken into account in the image reconstruction procedure. Multiple spiral interleaves as well as a multifrequency reconstruction algorithm are used to decrease blurring of off-resonance spins. Both phantom and in vivo images demonstrate the performance of the resulting pulse sequences.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.