Abstract

Spin lock (SL) imaging technique, generating T1 rho-weighted images, was applied to the differentiation of hepatic haemangiomas from metastatic focal liver lesions. 17 haemangiomas and 16 metastases in 32 patients were imaged at the field-strength of 0.1 T using a multiple slice SL technique and a conventional gradient-echo (GRE) sequence with identical timing parametres. Spin lock effects of the hepatic lesions and different abdominal tissues were calculated. Images with adequate coverage of the liver and of good quality with few motion induced artefacts were acquired. A definite, statistically significant, difference was found between the SL-effects of hepatic haemangiomas and a liver metastases. Haemangiomas showed an SL effect of 46.6 +/- 3.4% and metastases of 56.2 +/- 5.8% (mean +/- SD, p < 0.0001). The multiple slice SL technique showed potential in distinguishing haemangiomas from metastatic liver lesions and should be considered as an alternative to the conventional T2 and magnetization transfer (MT) based methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.