Abstract
Multiphoton microscopy has evolved into a powerful bioimaging tool in three dimensions. However, the ability to image biological specimens in-depth can be hindered by sample spherical aberration and scattering. These two phenomena can result in the degradation of image resolution and the loss of detected multiphoton signal. In this work, we use the correction collar (for cover glass thickness) associated with a water immersion objective in an attempt to improve multiphoton imaging. In the two samples we examined (human skin and rat tail tendon), we found that while the improvement in image resolution was not visible qualitatively, the measured axial fluorescence or second harmonic generation signal profiles indicate that the use of the correction collar can help to improve the detected multiphoton signals. The maximum increases are 36% and 57% for the skin (sulforhodamine B fluorescence) and tendon (second harmonic generation) specimens, respectively. Our result shows that for in-depth multiphoton imaging, the correction collar may be used to correct for spherical aberration. However, each tissue type needs to be examined to determine the optimal correction collar setting to be used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.