Spectrofluorometric in situ analysis of polycyclic aromatic compounds after separation on acetylated cellulose layers. I. Qualitative and quantitative evaluation The evaluation of polycyclic aromatic compounds, separated by thin-layer chromatography on acetylated cellulose, is described. In situ measurements of the fluorescence, excitation and emission spectra on thin-layer plates by means of a spectrofluorometer allow the identification as well as the determination of most of these substances. The limit of detection depends on the intensity of fluorescence of the aromatic compounds, and was estimated to be 0.1 to 0.001 μg on average. The error in the reproducibility in spotting the sample on the thin-layer plate is in the order of 10%. Standard curves show a fairly good linearity up to 0.2 μg of the substance. The results obtained from the standard curves, corrected by an external standard, are independent of the adjustment of the instrument.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.