Abstract

The intra-pulse spectroscopy based on room-temperature pulsed quantum cascade(QC) lasers was introduced. With a room-temperature pulsed QC laser centered at 1904 cm-1, the sample NO gas has been detected.The least-squared algorithm of baseline fitting and concentration retrieving method were introduced for the single line spectroscopy detection, and the gas concentration can be calculated with scan integration and the corresponding values from HITRAN04 database by direct-absorption technique, without calibration. The detection limit was obtained by analyzing the residual of the absorption spectroscopy as 34 ×10-6m.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.