Abstract
The intra-pulse spectroscopy based on room-temperature pulsed quantum cascade(QC) lasers was introduced. With a room-temperature pulsed QC laser centered at 1904 cm-1, the sample NO gas has been detected.The least-squared algorithm of baseline fitting and concentration retrieving method were introduced for the single line spectroscopy detection, and the gas concentration can be calculated with scan integration and the corresponding values from HITRAN04 database by direct-absorption technique, without calibration. The detection limit was obtained by analyzing the residual of the absorption spectroscopy as 34 ×10-6m.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.