In this work we evaluate the intensity and the spectral shape of the leakage radiation from the built-in beta and alpha sources of a Risø TL/OSL reader. LiF (TLD-100), fluorite-based pellets and Al2O3:C detectors were used in order to determine the dose rate delivered to the dosimeters when the sources are closed. The leakage spectra under both alpha and beta sources were registered with a CdTe semiconductor detector. The spectrum measured under the beta source shows the X-ray beam generated by the interaction of the beta particles with the lead used to shield the source. Besides, the 59.4 keV gamma ray from 241Am was registered under the alpha source. Dose rates from 50 to 100 μGy/h were obtained for the dosimeter positions in the turntable under the beta and the alpha sources with the luminescent dosimeters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.