Abstract

Development of a pulsed quantum cascade laser (QCL)-based spectroscopic trace-gas sensor for sub-part-per-million detection of nitric oxide (NO) and capable of monitoring other molecular species such as CO2, H2O, and NH3 in industrial combustion exhaust systems is reported. Rapid frequency modulation is applied to the QCL to minimize the influence of fluctuating non-selective absorption. A novel method utilizes only a few laser pulses within a single wavelength scan to probe an absorption spectrum at precisely selected optical frequencies. A high-temperature gas cell was used for laboratory evaluation of the NO sensor performance. A noise-equivalent sensitivity (1σ) of ∼ 100 ppb × m/\(\sqrt{Hz}\) at room temperature and ∼ 200 ppb × m/\(\sqrt{Hz}\) at 630 K was achieved by measuring the NO R(6.5) absorption doublet at 1900.075 cm−1.