Abstract

Optical remote sensing and Earth observation instruments rely on precise radiometric calibrations which are generally provided by the broadband emission from large-aperture integrating spheres. The link between the integrating sphere radiance and an SI-traceable radiance standard is made by spectroradiometer measurements. In this work, the calibration efforts of a Spectra Vista Corporation (SVC) HR-1024i spectroradiometer are presented to study how these enable radiance transfer measurements at the Calibration Home Base (CHB) for imaging spectrometers at the Remote Sensing Technology Institute (IMF) of the German Aerospace Center (DLR). The spectral and radiometric response calibrations of an SVC HR-1024i spectroradiometer are reported, as well as the measurements of non-linearity and its sensitivity to temperature changes and polarized light. This achieves radiance transfer measurements with the calibrated spectroradiometer with relative expanded uncertainties between 1% and 3% (k=2) over the wavelength range of 380 nm to 2500 nm, which are limited by the uncertainties of the applied radiance standard.