Abstract
Abstract We developed a spectropolarimetry system with a ferroelectric liquid crystal polarization modulator for the 60-cm Domeless Solar Telescope of Hida Observatory. The telescope is equipped with two oblique-reflection mirrors, that alter the incident polarization before it reaches the polarimeter. Furthermore, it is impossible to calibrate the polarimeter using unpolarized light because of the instrumental polarization. Here, we develop a procedure to obtain the parameters of both the instrumental polarization and for the calibration of the polarimeter from the same calibration data sets. We apply the new procedure to carry out high-sensitivity spectropolarimetry. It is now possible to measure the polarizations of various photospheric and chromospheric lines with the Domeless Solar Telescope.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.