Abstract

The express method for testing purity and stability of an indotricarbocyanine dye based photosensitizer for photodynamic therapy was developed. Storage of the hydrophilic photosensitizer under unfavorable conditions resulted in contamination with a hydrophobic dye. The absorption spectrum of the hydrophobic dye was characterized by a new band at 514 nm that was absent in the absorption spectrum of the photosensitizer. There was no spectral overlap between this absorption band and the absorbance of the photosensitizer. Therefore, presence of the band at 514 nm allowed us to detect contamination of the photosensitizer substance with the hydrophobic dye. It was established that the optimal photosensitizer concentration for purity testing is ca. 0.8 mmol/L, when as little as 0.6 wt. % of the impurity can be detected. The developed method is not time consuming and requires accessible equipment only. Data obtained with this method was verified with the aid of liquid chromatography-mass spectrometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.