Abstract

A method for measuring the groove depth of calibration gratings is proposed which is based on measuring the spectral dependence of the the zero-order reflection diffraction efficiency. The errors of the method are determined by three main factors: the shift of the maxima of the spectrum due to the wall slope of the grating grooves, the error in setting the wavelength of the spectrophotometer, and the divergence of the light beam in the setup. It is shown theoretically that the measurement error is in the range of 0.25–1%, depending on the fabrication technology of the grating and measuring equipment. The method was tested experimentally using commercial calibration gratings. The range of applicability of the method is discussed in terms of the geometrical parameters of the microstructure of reflection gratings and the characteristics of the spectrophotometer used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.