Abstract

Continuous monitoring of concentration is important in the optimal control of bioreactors. Spectrophotometry provides a simple, accurate, and rapid way for measuring cell concentration of unicellular microorganisms, based on either Beer's law or calibration curves prepared using standard solutions. However, this method is limited to a low range of microbial concentrations, because Beer's law deviates significantly at high concentrations. In the present investigation, based on experimental work, a new technique is posed to monitor the concentration of microbial cells as high as 100 g DW/L. this is achieved by using a mixture of known concentration as reference, rather than the "ideal blank" with zero concentration of analyte. As a result, a new equation is developed that, although applied here only to microbial concentration, in principle can be used for monitoring the concentration of any optically sensitive material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call