Abstract
A simple, rapid, and accurate spectrophotometric flow-injection method was developed for the determination of sorbic acid in wines based on the oxidation of sorbic acid with K2Cr2O7/H2SO4 at 1008C, followed by reaction of the resulting malonaldehyde with thiobarbituric acid, also at 1008C, to give a red product, the absorbance of which is measured at 532 nm. The method requires no prior distillation of the wine and allows a sampling rate of 40 per hour. The calibration graph was linear from 0 to at least 15 lg mL11 (the linear correlation coefficient was 0.9997), and the detection limit (signal-to-noise ratio, 3) was 0.14 lg mL11. The relative standard deviation of ten replicate determinations was 1.58%. The method was applied successfully to the determination of sorbic acid in red and white wines. q 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 299–303, 1999
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.