Abstract
A spot test was implemented in a flow-injection system for the spectrophotometric determination of zinc in digests of plant materials. It is based on the influence of Zn2+ on the oxidation rate of 1-naphthylethylenediamine (NED) by hexacyanoferrate(III) under acidic conditions. In order to control the precipitate formation and to maintain the resulting suspension, a micellar medium was established by adding Triton X-100. The proposed system handles about 65 samples per hour, meaning 72μg NED and 9.0mg K3[Fe(CN)6] per determination. Baseline drift is usually <0.01 absorbance per hour and the analytical signals for 0.5–2.5mgl−1 Zn range within ca. 0.07–0.45 absorbance. Linearity of the analytical curve is fair (r>0.999, n=6) and detection limit was estimated as 0.2mgl−1 Zn. Results are precise (R.S.D.<1%, n=10) and in agreement with flame atomic absorption spectrometry and with certified values of standard reference materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.