Abstract

A rapid, relatively sensitive, and low-cost method for the determination of water-soluble urea content in dermatological therapy products and cosmetics is proposed using a new spectrophotometric assay with water as the only extraction solvent. Spectrophotometric methods involve addition of a known excess of bromate to urea in an acid medium, followed by the determination of residual bromine and chlorine reacting with methyl orange and measurement of absorbance at 505 nm. The absorbance increases linearly with urea concentration (r = 0.9998). The systems obey Beer's law for 6 - 90 microg ml(-1). The calculated apparent molar absorbance values are found to be 4.537 x 10(3) dm(3) mol(-1) cm(-1) and the Sandell's sensitivity is 0.013 microg cm(-2). The variables affecting the rate of the reaction were investigated. The relative standard deviation for five-replication determination of 60 microg ml(-1) urea was 2.1% and the detection limit of the method is 0.34 ng ml(-1).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.