Abstract

An approach to automation, in routine control, for the determination of anionic surfactants in river water and treatment plant water in the concentration range 0.04–3.5 µg ml–1 using a continuous solvent extraction process with flow injection analysis, is described. The method is based on an ion-pair extraction reaction with Methylene Blue in chloroform. The separation of both phases is accomplished by means of a membrane phase separator and detection is based on injection of the coloured organic phase into a chloroform stream which passes to the detector. Methanol was added to the organic phase in order to enhance the extraction process and to ensure efficient extraction of the various anionic surfactants. A number of anionic species and non-ionic surfactants have been tested as possible interferents. The results obtained for samples of river water show good agreement with those obtained with the Methylene Blue batch method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.