Abstract
The determination of ethanol in whole blood without pretreatment using spectrotometric detection with an immobilized enzyme reactor coupled to an on-line dialyser is described. Optimum working conditions are given. The effects of the injection volume and the flow-rate on the peak height using the dialyser were investigated. A good linear calibration graph over the ethanol concentration range 3–40 μg ml−1 was observed; these concentrations correspond to 0.3−4.0 g of ethanol per 1000 g of whole blood prior to dilution. For comparison, whole blood samples were analysed by gas chromatography with direct injection. The effect of temperature on the peak height was also studied in a system without the dialyser.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.