Abstract

This paper focuses on a revision of the phenolphthalin method for cyanide analysis, with the intent of producing a robust and sensitive spectrophotometric method. Limitations of the phenolphthalin chemistry were overcome by the addition of EDTA to the reagent. The revised reagent was found to have suitably fast kinetics, a linear dynamic range of 0.01–3.0 ppm cyanide, and a detection limit of 5 ppb. The method was tested for interferences and applied to the determination of cyanide in environmental waters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.