Abstract

The distinction between the terms photometric accuracy, photometric linearity, and adherence to Beer's Law is discussed. To describe the last, a new term “analytical linearity” is suggested. Some instrument response functions which exhibit acceptable analytical linearity are shown, nevertheless, to be photometrically inaccurate. Methods of measuring true photometric accuracy, the search for absolute photometric standards, and the ultimate limitations of such standards are reviewed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.