Abstract

The spectrometer uses an array of 50 pyroelectric detectors etched from a PVF2 sheet in the precise positions for the various lines in the P and R branches of the CO2 spectrum. The detectors are accessed in sequence to give a bar chart display of the spectrum. Compared to the conventional spectrum analyser (using a phosphor screen for CW or a graphite screen for pulsed lasers) the instrument gives a substantial advantage in sensitivity with a noise equivalent power of 100 mu W for CW or 200 nJ for pulsed. The dynamic range is 3000:1, enabling the detection of weak satellite lines, while a digital store allows monitoring of each shot of an inconsistent but low repetition rate laser.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.