Abstract

Trace amounts of beryllium has been determined by spectrofluorimetric method that used morin as fluorimetric reagent. Beryllium gives a highly fluorescent complex with morin. The excitation wavelength of morin and Be-morin complex were 410 and 430. The fluorescence spectra of morin and Be-morin complex were overlaped in excitation wavelength of 430nm. A method based on mean centering of ratio spectra has been performed to remove the interference caused by morin as it overlaps with the Be-morin spectra. The linear range of beryllium concentration is in 0.2-200ppb range. The parameters of detection limit and RSD were 0.18ppb and 4.6% respectively. This method was used for determination of beryllium in copper-beryllium alloy as a real sample. In determination of Be(II), the interference by Cu(II) was very serious, which was eliminated by adding triethanolamine.