Abstract

Infrared radiometers for irradiance measurement have been developed at the National Institute of Standards and Technology (NIST). These high performance irradiance meters are used to realize and maintain the spectral irradiance responsivity scale between 1000 nm and 5000 nm. They are also working standards that disseminate the infrared irradiance responsivity scale to other institutions and facilities. Both design considerations and responsivity scale transfer to the irradiance meters are discussed. The radiometers were calibrated for spectral irradiance responsivity on the new NIST Infrared Facility for Spectral Irradiance and Radiance Responsivity Calibrations using Uniform Sources (IR-SIRCUS). The spectral irradiance responsivity calibrations described are derived from absolute cryogenic radiometers.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.