Abstract

A method of measuring the absolute spectral irradiance of quartz-halogen-tungsten lamps is described, based on the known responsivity of a filter radiometer, the components of which are separately characterized. The characterization is described for the wide wavelength range essential for deriving the spectrum of a lamp, from 260 nm to 950 nm. Novel methods of interpolation and measurement are implemented for the spectral responsivity of the filter radiometer. The combined standard uncertainty of spectral irradiance measurements is less than 1.4 parts in 102 from 290 nm to 320 nm (ultraviolet B) and 4 parts in 103 from 440 nm to 900 nm (visible to near-infrared). As an example, the derived spectral irradiances of two lamps measured at the Helsinki University of Technology (HUT, Finland) are presented and compared with the measurement results of the National Institute of Standards and Technology (NIST, USA) and the Physikalisch-Technische Bundesanstalt (PTB, Germany). The comparisons indicate that the HUT spectral irradiance scale is between those of the NIST and the PTB in the wavelength range 290 nm to 900 nm. The long-term reproducibility of the spectral irradiance measurements is also presented. Over a period of two years, the reproducibility appears to be better than 1 part in 102.